Seite 1 - Digital Thermometer

OMEGAHH505Digital Thermometer

Seite 2 - Servicing Europe:

8101112Hi Limit value Recall display modePress Hi RECALL key to display the Hi set Limit value on second display.Press Hi RECALL key again to exit th

Seite 3 - SAFETY INFORMATION

9WARNINGTo avoid possible electrical shock, disconnect the thermocoupleconnectors from the rmometer before removing the cover.Battery ReplacementPower

Seite 4 - SPECIFICATIONS

10TEMPERATURE VS RESISTANCE TABLE(ITS90)°C Pt385 Pt3926-200°C 18.521W 16.996W-100°C 60.256W 59.479W0°C 100.000W 100.000W100°C 138.505W 139.272W200°C 1

Seite 5 - ENVIRONMENTAL

1HH 504 / HH 505 CALIBRATION PROCEDURENote: The following calibration procedure should perform only by qualified technicianswho have access to the ite

Seite 6 - OPERATING INSTRUCTIONS

WARRANTY / DISCLAIMEROMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 monthsfrom date

Seite 7 - MIN MAX with Time record Mode

Where Do I Find Everything I Need forProcess Measurement and Control?OMEGA...Of Course!TEMPERATUREþ Thermocouple, RTD & ThermistorProbes, Connecto

Seite 8 - REL Relative mode

OMEGAnetSM On-Line Servicehttp://www.omega.comInternet [email protected] North America:USA: ISO 9001 Certified Canada:One Omega Drive, Box

Seite 9

1The symbol on the instrument indicates that the operator must refer to anexplanation in this manual.WARNINGTo avoid electrical shock, do not use th

Seite 10

2SPECIFICATIONSELECTRICALTemperature Scale: Celsius or Fahrenheit user-selectable.Measurement Range:Pt385(100W ) -200°C to 800°C, (-328°F to 1472°F)Pt

Seite 11 - P/N : 17650-0505

3ENVIRONMENTALAmbient Operating Ranges: 0°C to 50°C (32°F to 122°F) <80% R.H.Storage Temperature: -20°C to 60°C (-4°F to 140°F) <70% R.H.GENERAL

Seite 12

OPERATING INSTRUCTIONS Power SwitchThe key turns the thermometer on or off. In the data SET mode, can notpower off, must leave data SET mode

Seite 13 - 383.00W CALIBRATOR

556Pt385/Pt3926 T1 Input RTD Probe SelectThe Pt385/Pt3926 key switch the T1 input selects the Pt385 or Pt3926 RTDprobe as input.When the thermometer

Seite 14 - RETURN REQUESTS / INQUIRIES

6REL Relative modePressing REL key to enter the Relative mode, zero the display, and storethe displayed Reading as a reference value and annunciator R

Seite 15 - OMEGA...Of Course!

79SET mode (Relative value set, Time set and Hi/Lo Limits value set)1. Press SET key to enter Relative valuse SET mode (Press ENTER key canescape rela