Omega Engineering HH504 Bedienungsanleitung

Stöbern Sie online oder laden Sie Bedienungsanleitung nach Messen, Testen & Steuern Omega Engineering HH504 herunter. Omega Engineering HH504 User's Manual Benutzerhandbuch

  • Herunterladen
  • Zu meinen Handbüchern hinzufügen
  • Drucken

Inhaltsverzeichnis

Seite 1 - Digital Thermometer

OMEGAHH504Digital Thermometer

Seite 2 - Servicing Europe:

81011124. When the thermometer is turned on. The Relative set value and Hi/Lo Limitsset value that was in use when thermometer was last turned off set

Seite 3 - SAFETY INFORMATION

9RTD PROBE CONNECTIONP/N : 17650-0505RW1=RW2=RW3RW= WIRE RESISTANCEG+-RTD100WPt385(Pt3926)WARNINGTo avoid possible electrical shock, disconnect the th

Seite 4 - SPECIFICATIONS

10TEMPERATURE VS RESISTANCE TABLE(ITS90)°C Pt385 Pt3926-200°C 18.521W 16.996W-100°C 60.256W 59.479W0°C 100.000W 100.000W100°C 138.505W 139.272W200°C 1

Seite 5 - ENVIRONMENTAL

1HH 504 / HH 505 CALIBRATION PROCEDURENote: The following calibration procedure should perform only by qualified technicianswho have access to the ite

Seite 6 - OPERATING INSTRUCTIONS

WARRANTY / DISCLAIMEROMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 monthsfrom date

Seite 7

Where Do I Find Everything I Need forProcess Measurement and Control?OMEGA...Of Course!TEMPERATUREþ Thermocouple, RTD & ThermistorProbes, Connecto

Seite 8

OMEGAnetSM On-Line Servicehttp://www.omega.comInternet [email protected] North America:USA: ISO 9001 Certified Canada:One Omega Drive, Box

Seite 9

1WARNINGTo avoid electrical shock, do not use this instrument when workingvoltages at the measurement surface over 24V AC or DC.WARNINGTo avoid damage

Seite 10

2SPECIFICATIONSELECTRICALTemperature Scale: Celsius or Fahrenheit user-selectable.Measurement Range:Pt385(100W ) -200°C to 800°C, (-328°F to 1472°F)Pt

Seite 11 - Battery Replacement

3ENVIRONMENTALAmbient Operating Ranges: 0°C to 50°C (32°F to 122°F) <80% R.H.Storage Temperature: -20°C to 60°C (-4°F to 140°F) <70% R.H.GENERAL

Seite 12

231OPERATING INSTRUCTIONS Power SwitchThe key turns the thermometer on or off. In the data SET mode, can notpower off, must leave data SET mo

Seite 13 - 383.00W CALIBRATOR

5465T1/ T2/T1-T2 Main display Input SelectionThe input selection indicates which input is selected for main display; T1probe, T2 probe or the differen

Seite 14 - RETURN REQUESTS / INQUIRIES

67The true average of all the reading taken over at least 22 hours preiod canbe displayed. If 22 hours is exceeded, new averages are no longercalculat

Seite 15 - OMEGA...Of Course!

89°C/°F Selecting the Temperature Scale (Second display)Readings the second displayed in either degrees Celsius(°C) or degreesFahrenheit(°F). When the